Anterolateral S1 screw malposition detected with intraoperative neurophysiological monitoring during posterior lumbosacral fusion
-
Received: ,
Accepted: ,
Abstract
Background:
The standard of care is to utilize intraoperative neurophysiological monitoring (IOM) of triggered electromyography (tEMG) during posterior lumbosacral instrumented-fusion surgery. IOM should theoretically signal misplacement of S1 screws into the neural L5–S1 foramen or spinal canal, utilizing screw stimulation, and recording of the lower limb muscles and the anal sphincter. Here, we evaluated when and whether anterolateral S1 screw malposition could be detected by IOM/tEMG during open posterior lumbosacral instrumented fusion surgery.Methods:
tEMG, somatosensory-evoked potential (SSEP), and transcranial electrical motor-evoked potential (TcMEP) data were retrospectively reviewed from 2015 to 2017 during open posterior lumbosacral instrumented fusions. We utilized screw stimulation alert thresholds of <14 mA (tEMG) and recorded from the lower extremity muscles and anal sphincter. Furthermore, all patients underwent routine postoperative computed tomography (CT) scans to confirm the screw location.Results:
There were 106 S1 screws placed in 54 patients: 52 bilateral and 2 unilateral. In 6 patients (11.1%), 7 screws (6.6%) registered at low tEMG thresholds. In 1 patient, the postoperative CT scan documented external malposition of the screw despite no intraoperative IOM/tEMG alert. When S1 misplaced screws were stimulated, the most sensitive muscle was the tibialis anterior; the sensitivity of the IOM/tEMG was 87.5%, the specificity was 97.9%, the positive predictive value was 77.8%, and the negative predictive value was 98.9%. TcMEP and SSEP did not change during any of the operations. Notably, no patient developed a new neurological deficit.Conclusion:
Anterolateral S1 screw malposition can be detected accurately utilizing IOM/tEMG stimulation of screws. When alerts occur, they can largely be corrected by partially backing out the screw (e.g., a few turns) and/ or changing the screw trajectory.Keywords
INTRODUCTION
Posterior lumbosacral fixation with screws is often the treatment of choice for addressing multiple spinal pathologies. Nevertheless, the malposition of screws risks major vessel[3] and/or neural structural[2] injuries.
Intraoperative neurophysiological monitoring (IOM) techniques[8] (triggered electromyography [tEMG], somatosensory-evoked potentials [SSEP], and transcranial electrical motor-evoked potential [TcMEP]), used along with fluoroscopy[6] and/or navigation assistance[4] increase the accuracy of screw placement and patient safety during lumbosacral spinal instrumented fusions.
Here, we specifically focused on the utility of IOM/tEMG for detecting anterolateral S1 screw malposition during posterior lumbosacral fusion surgeries.
MATERIALS AND METHODS
We retrospectively reviewed IOM/tEMG (e.g., along with SSEP and TcMEP). Data from patients undergoing open posterior lumbosacral pedicle/screw instrumented fusions (2015–2017) for L5–S1 instability, spinal canal, and/or foraminal stenosis.
To avoid interference with IOM, anesthesia utilized total intravenous anesthesia without neuromuscular blockade.
EMG recordings and stimulation
EMG recordings were taken using 13-mm subdermal needle electrodes inserted into muscle groups innervated by the nerve roots most at risk of injury from screw placement. These muscles include bilateral quadriceps femoris, tibialis anterior, abductor hallucis, gastrocnemius medial, and anal sphincter. They were filtered at 30–3000 Hz.
Pedicle screw stimulation of S1 was performed by the cathodal constant current with 200 µS of duration at 2 Hz, and the response recorded from lumbosacral nerve roots, including bilateral anal sphincter monitoring. Response from S1 screw stimulation was utilized to help determine anterolateral S1 screw malposition. In addition, differences in the latency of L5 nerve root activation helped confirm superior foraminal S1 screw malposition or anterolateral S1 malposition. After backing out the screw and/or changing the screw trajectory, re-stimulation of screws documented whether there was an anterolateral malpositioning of the screw. Lateral X-rays also helped to rule out L5-S1 foraminal misplacement. Further, 1-day postoperative computed tomography (CT) scans helped confirm the final position of the lumbosacral S1 screws.
Patient population
There were 106 S1 screws placed during open posterior lumbar instrumented fusions in 54 patients (e.g. 52 bilateral and 2 unilateral). All patients had preoperative lumbar magnetic resonance imaging studies confirming: (a) the pathology: stenosis versus degenerative spondylolisthesis versus spondylolysis and (b) to document the location of the L5 nerve root within the L5–S1 foramen [Figures 1 and 2].


RESULTS
Malpositioned screws with IOM/tEMG alerts
In 6 patients (11.1%), there were 7 positioned screws (6.6%) (e.g., IOM/tEMG confirmed a low threshold obtained) (<14 mA) [Figures 3a and 3b]. Adjustments were made by back turning of 4 S1 screws. Repeat stimulation of screws was then performed and documented significant sufficiently increased threshold (>14 mA), consistent with adequate screw placement [Figures 3c and 3d], and confirmed by postoperative computed tomography scan [Figure 4]. Other 3 S1 screws with a low threshold obtained [Figure 5a] required repositioning (e.g., medial trajectory adjustments at the same entry point), and repeat stimulation was performed and confirmed correct screw placement [Figure 5b], with a postoperative image demonstrated adequate position of the screw [Figure 6].




CT study of the malpositioned screw without IOM/tEMG alerts
On the postoperative CT scan, only 1 asymptomatic patient who had no intraoperative IOM/tEMG changes had malpositioning of the S1 screw.
Greatest sensitivity of anterior tibialis IOM/tEMG recording for S1 screw malpositioning
When placing S1 screws, the anterior tibialis muscle most readily detected screw malpositioning [Figures 3a, b and 5]. The overall sensitivity of S1 IOM/tEMG was 87.5%; specificity was 97.9%, positive predictive value was 77.8%, and negative predictive value was 98.9%.
No significant SSEP/TcMEP Changes
Notably, TcMEP and SSEP did not show any significant changes throughout all the surgeries, and no patient developed a new neurological deficit.
DISCUSSION
The L5 nerve root injury arising from a S1 outwardly misplaced screw is a well-known complication in spine surgery. Inoue et al.[5] published two symptomatic cases with S1 screw malpositioning. Placing S1 sacral screws is challenging and carries the risks of major vascular/neural injuries.[2,3] To avoid these injuries, IOM/tEMG has been routinely adopted in spine surgery during pedicle screw placement/lumbosacral fusions.[1] Here, we evaluated the efficacy/accuracy of IOM/tEMG in detecting anterolateral S1 screw malpositioning during open lumbosacral instrumented spinal fusions. We can consider a breach to be likely if the tEMG response was <14 mA.
It is strongly recommended that sacral screws use an entry point just lateral to the inferior boundary of the L5–S1 facet joint, as described by Weinstein and Magerl. In contrast, Roy-Camille described a medial entry point which has some advantages over traditional techniques.[7] (e.g. leading to screw placement more laterally avoiding the spinal canal at S1) [Figure 7].

Of interest, the mixed posterior tibial nerve SSEP did not result in any significant changes despite S1 screw misplacement. Further, TcMEP showed no significant changes.
CONCLUSION
IOM/tEMG accurately detected anterolateral S1 screw malpositioning occurring during open lumbosacral instrumented fusions. These alerts allow for (a) back turning of screws and/or (b) redirecting screw trajectories intraoperatively to avoid neural compromise.
Declaration of patient consent
Patient’s consent not required as patient’s identity is not disclosed or compromised.Financial support and sponsorship
Nil.Conflicts of interest
There are no conflicts of interest.REFERENCES
- Spine (Phila Pa 1976). 1994;19:2780-6.Stimulus-evoked EMG monitoring during transpedicular lumbosacral spine instrumentation. Initial clinical results.
- [Google Scholar]
- Orthop Traumatol Surg Res. 2014;100:S85-90.Neurological risks in scheduled spinal surgery.
- [Google Scholar]
- J Vasc Surg. 2011;53:407-13.Iatrogenic arterial injuries of spine and orthopedic operations.
- [Google Scholar]
- Neurosurg Focus. 2014;36:E2.Accuracy of intraoperative computed tomography image-guided surgery in placing pedicle and pelvic screws for primary versus revision spine surgery.
- [Google Scholar]
- Eur Spine J. 2013;22:S461-5.L5 spinal nerve injury caused by misplacement of outwardly-inserted S1 pedicle screws.
- [Google Scholar]
- Clin Spine Surg. 2018;31:108-11.Intraoperative radiographic technique for visualization of bicortical or tricortical anteromedial sacral screw placement.
- [Google Scholar]
- J Orthop Surg Res. 2014;9:22.Evidence-based support for S1 transpedicular screw entry point modification.
- [Google Scholar]
- Phys Med Rehabil Clin N Am. 2004;15:85-105.Electrophysiologic intraoperative monitoring for spine procedures.
- [Google Scholar]